Lw-scientific Digital ZipSpin Bedienungsanleitung

Stöbern Sie online oder laden Sie Bedienungsanleitung nach Optik Lw-scientific Digital ZipSpin herunter. LW Scientific Digital ZipSpin User Manual Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 8
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 0
P. 770.270.1394 F. 770.270.2389 865 Marathon Parkway Lawrenceville GA 30046
Instruction Manual
Model pictured:
Digital ZipSpin Centrifuge
24 Volt Digital ZipSpin24 Volt Digital ZipSpin
The LW Scientific Digital ZipSpin is loaded with user-friendly features such as digital read out of time and speed, digital
calibration, imbalance detection, and last cycle run recall. Designed with robust construction for long life, the Digital
ZipSpin is well-suited to accomplish blood and urine separations, micro filtrations, and DNA preparations. Since more and
more tests are done with smaller sample sizes, this design includes two interchangeable rotors: an 8-place closed rotor for
1.5/2.0ml micro-tubes and a PCR strip rotor (16 places) for 0.2ml micro-tubes. Also included are 2 sets of adapters for
0.4/0.5ml micro-tubes and 0.2ml micro-tubes. The small footprint and quiet operation make it the perfect addition to any
clinic or lab.
Please read and complete the warranty form online at LWScientific.com/warranty_form. The warranty form
documents your purchase. Failure to fill out the warranty form may void any warranty claims on the unit.
The Digital ZipSpin centrifuge is designed to be safe under the following conditions when operated with necessary
precautions:
• For indoor use only.
• Ambient temperature of 4°C to 65°C.
• Main supply voltage fluctuations not to exceed +/- 10% of the nominal voltage.
• Position at least 20 cm from walls, 20 cm from ceiling and 20 cm between other products.
• Make sure to always install the centrifuge on stable flat surfaces only & place it near the electrical outlet.
• Make sure to always load the rotor symmetrically every time. Each tube should be counterbalanced by another tube.
• Make sure to always use only specified micro centrifuge tubes made from plastic and designed to withstand
centrifugal forces of at least 2000 g.
• Do not use the centrifuge in any manner not specified in these instructions.
• Do not operate the centrifuge without the rotor being properly attached to the shaft. Re-check by trying to pull it up
manually, especially after rotor assembly.
• Do not move the centrifuge while the rotor is spinning.
• Do not fill tubes while they are in the rotor, liquid spillage may harm the unit.
• Do not put hands in the rotor unless the rotor has completely stopped.
• Do not centrifuge flammable, explosive or corrosive materials.
• Do not use solvents or flammables near this or other electrical equipment.
• In case of the unit dropping from the table please check the casing for any cracks/damage.
• The centrifuge and its components are non-autoclavable.
• Speeds for tube rotor are accurate to +/- 5%
• Speeds for PCR rotor are accurate to +/- 12%
A Lid
B Rotor
C Timer Knob
D Speed Knob
E Display
F Release Button
G Safety Switch
H ON/OFF Switch
I Non-Slip Rubber Base
J Power Socket
A
G
B
F
D
C
E
H
I
J
Seitenansicht 0
1 2 3 4 5 6 7 8

Inhaltsverzeichnis

Seite 1 - Instruction Manual

P. 770.270.1394 F. 770.270.2389 865 Marathon Parkway Lawrenceville GA 30046Instruction ManualModel pictured:Digital ZipSpin Centrifuge24 Volt Digital

Seite 2 - Installation

OperationSwitch ‘On’ the centrifuge from rear side so that display will show you the readings.Note : ‘En’ Symbol indicates that Imbalance detection fe

Seite 3 - IMPORTANT:

P. 770.270.1394 F. 770.270.2389 865 Marathon Parkway Lawrenceville GA 30046Balancing RotorTo spin 8 samples, fill 8 test tubes equally. Then insert t

Seite 4 - Changing the Rotor

Interchangeable Rotor8-Place Rotor8 x 0.2 ,0.4 ,0.5 ,1.5 ,2 mlPCR Rotor16 x 0.2 mlChanging the RotorInsert the L-arm hex wrench into the marked hole o

Seite 5

P. 770.270.1394 F. 770.270.2389 865 Marathon Parkway Lawrenceville GA 30045Time & Speed Settings and DisplayThis unit is equipped with a 0-25 minu

Seite 6 - Imbalance Detection

Time & Speed Settings and Display Cont.During the operation of centrifuge, the time and speed will be displayed alternatively accompanied by speed

Seite 7 - Calibration

P. 770.270.1394 F. 770.270.2389 865 Marathon Parkway Lawrenceville GA 30046CalibrationYour centrifuge comes factory-calibrated for accurate speeds wit

Seite 8 - Operating Conditions

P. 770.270.1394 F. 770.270.2389 865 Marathon Parkway Lawrenceville GA 30046Operating ConditionsMKT-7.5.3-L-183 | Rev 0The Digital ZipSpin centrifuge i

Kommentare zu diesen Handbüchern

Keine Kommentare